Robert A. Lookstein, M.D, et al
BACKGROUND |
Standard percutaneous transluminal angioplasty is the current recommended treatment for dysfunctional hemodialysis fistulas, yet long-term outcomes of this treatment are poor. Drug-coated balloons delivering the antirestenotic agent paclitaxel may improve outcomes
METHODS |
In this prospective, single-blinded, 1:1 randomized trial, we enrolled 330 participants at 29 international sites. Patients with new or restenotic lesions in native upper-extremity arteriovenous fistulas were eligible for participation. After successful high-pressure percutaneous transluminal angioplasty, participants were randomly assigned to receive treatment with a drug-coated balloon or a standard balloon. The primary effectiveness end point was target-lesion primary patency, defined as freedom from clinically driven target-lesion revascularization or accesscircuit thrombosis during the 6 months after the index procedure. The primary safety end point, serious adverse events involving the arteriovenous access circuit within 30 days, was assessed in a noninferiority analysis (margin of noninferiority percentage points 7.5).The primary analyses included all participants with available end-point data. Additional sensitivity analyses were p erformed to assessthe effect of missing data
RESULTS |
A total of 330 participants underwent randomization; 170 were assigned to receive treatment with a drug-coated balloon, and 160 were assigned to receive treatment with a standard balloon. During the 6 months after the index procedure, targetlesion primary patency was maintained more often in participants who had been treated with a drug-coated balloon than in those who had been treated with a standard balloon (82.2% [125 of 152] vs. 59.5% [88 of 148]; difference in risk, 22.8 percentage points; 95% confidence interval [CI], 12.8 to 32.8; P<0.001). Drugcoated balloons were noninferior to standard balloons with respect to the primary safety end point (4.2% [7 of 166] and 4.4% [7 of 158], respectively; difference in risk, −0.2 percentage points; 95% CI, −5.5 to 5.0; P = 0.002 for noninferiority). Sensitivity analyses confirmed the results of the primary analyses
CONCLUSIONS |
Drug-coated balloon angioplasty was superior to standard angioplasty for the treatment of stenotic lesions in dysfunctional hemodialysis arteriovenous fistulas during the 6 months after the procedure and was noninferior with respect to access circuit– related serious adverse events within 30 days. (Funded by Medtronic; IN.PACT AV Access Study ClinicalTrials.gov number, NCT030414
دیدگاه خود را بنویسید